ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-15
2558- The instrument is busy.
Cause: An instrument is busy. The requested operation cannot be executed.
Action: The instrument is busy. The requested operation cannot be done at this
time.
Action: Wait until the instrument is idle and
try the operation again.
The requested operation cannot be executed.
2559- SPC directory (name) does not exist.
Cause: Files were selected for SPC reporting from a directory which no longer
exists.
Action: If the directory was deleted, renamed, or moved in error, restore the
dire
ctory.
Action: Select a different set of files for SPC reporting.
2560- File (filename) cannot be created or opened. It has an unrecognized extension.
Cause: A file name with an extension unrecognizable by the software was
typed into a file selector.
Action: Type in a different file name. The extensions used by the software are:
Action: If the named directory was accidentally renamed or move, restore the
dire
ctory.
Action: If the file reference is incorrect, re
move the file reference.
SMP - Sample File
ANC - Analysis Conditions
DEG - Degas Conditions
ADP - Adsorptive Properties
RPO - Report Options