ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-35
6218- SmartVac is not in a valid state to check degas.
Cause: You clicked Check to check degassing but the sample was not in an
appropriate state. You can only check the degassing operation after the
vacuum setpoint has been attained, or during a temperature ramp or
hold.
Action: Wait until an appropriate time during t
he degassing operation and
check the degassing operation again.
6219- The file name (name) has already been used.
Cause: You entered a file name that is already in use.
Action: Enter a unique file name.
6220- There are currently no available ports to start a degas operation.
Cause: All degas ports are in use.
Action: Wait until the degas operation is complete on a port, then another degas
op
eration may be started.
6221- An evacuation cannot be started because all degas ports are currently in use.
Cause: All degas ports are in use.
Action: Wait until the degas operation is complete on a port, then
an evacuation
can be started on the port.
6222- At least one port must be selected before an evacuation can be started.
Cause: You clicked Start on the Evacuate Degas dialog without first selecting
a port.
Action: Select the port you wish to evacuate, then
select Start.