Appendix B ASAP 2460 Operator’s Manual
B-36 246-42800-01 - Dec 2013
6223- There is no Nitrogen attached to the unit.
Cause: A calibration requiring nitrogen was attempted but the software does
not recognize that nitrogen is attached.
Action: Ensure that a nitrogen gas bottle is installed at one of the analysis ports,
then Select Unit
> Unit configuration and enter N2 for the appropriate
valve. Refer to Unit Configuration, page 4-2
9.
6226- An analysis cannot be started because all ports are currently is use.
Cause: You tried to start an analysis while all ports are in use.
Action: Wait until the analysis completes on a port, the
n another analysis can
be started.
6227- File (name) cannot be analyzed. It is currently being edited.
Cause: You tried to start an analysis using a sample information file that is cur-
rently being edited.
Action: Close the edit session on the sample information file
, then the file can
be used for analysis
.
6228- Analysis gas in sample file does not match analysis gas in other active
sa
mples.
Cause: You tried to start an analysis using a sample file that contains an anal-
ysis gas other than the gas currently being us
ed. All current analyses
must use the same analysis gas.
Action A: Change the gas in the sample file to us
e the current analysis gas.
Action B: Wait until the current analyses are
complete, then run the sample.
6229- This file already selected for the analysis.
Cause: You selected a sample file that is already being used for an analysis.