ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-37
6230- Dosing method does not match analysis type.
Cause: You tried to perform a nitrogen or nitrogen-like analysis with From
last port selected as the dosing method.
Action: Change the Dosing Method to Normal in
the Adsorptive Properties
dialog, then start the analysis.
6231- Dosing method does not match analysis type.
Cause: You tried to perform a Krypton analysis with Normal selected as the
dosing method.
Action: Choose F
rom last port for Dosing Method in the Adsorptive Proper-
ties dialog, then start the analysis.
6232- Po gas in sample file does not match Po gas in other active samples.
Cause: You tried to start an analysis using a sample file that contains a Po gas
other than the Po gas currently being used. All current analyses must
use the same Po gas.
Action A: Change the gas in the sample file to use
the current Po gas.
Action B: Wait until the current analyses are complet
e, then run the sample.
Action: Select another sample information file.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals