EasyManua.ls Logo

Minebea Intec X3 Process Indicator PR 5410 - Page 154

Minebea Intec X3 Process Indicator PR 5410
340 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
— Test mode Selection for display of the deviation from the
test value: <Absolute>, relative
— W&M Selection legal-for-trade mode: <none>, OIML
(impossible when [Range mode] "Multi-
interval" has been selected or Max has more
than 3 decimals), NSC, NTEP
— Standstill time
Input of the standstill period: 0.01 s<0.50
s>2.0 s (The range depends on the measuring
time.)
— Standstill range
Input of the standstill range: 0.00 d<1.00
d>10.00 d
— Tare timeout Input of timeout when there is no standstill: 0.1
s<2.5 s>25 s
— Zeroset range ± Range of zero point when there is no
standstill. Input: 0.00 d<50.00 d>10000.00
d
— Zerotrack range
Input of the zerotrack range: 0.00 d<0.25
d>10000.00 d
— Zerotrack step
Input of the zerotrack step: 0.00 d<0.25
d>10.00 d
— Zerotrack time
Input zerotrack time: <0.0 s>25 s
— Overload Input of the weighing range above the
maximum load (Max) without error message:
0999999 d
— Minimum weight
Input of the minimum load: 0 d<50
d>999999 d
— Range mode Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.14.15.2 and 7.14.15.3.
— Range limit 1 Input of the limit 1: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— Range limit 2 Input of the limit 2: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— View (when CAL-switch is closed)
— Max Display only
— Scale interval Display only
— Dead load at Display only
— Max at Display only
— Calibrated at Display only
— Sensitivity (µV/d) Display only
— Param Items as for [Param] (display only)
— Setup Setup of the xBPI scale
— Calibration Calibration of xBPI scale
— Dead load Dead load/preload
— Set Set dead load/preload: Accept, ResError, Abort
— Delete Clear dead load/preload: Accept, ResError,
Abort
7.13.7.2 "xBPI scale" weighing point
X3 Process Indicator PR 5410 7 Commissioning
Minebea Intec EN-152