No. 99MBG518A
D-12
PART D : 2 Method to Test One Point
2.7.2 Executing the Test
Execute the test with the specimen for obtaining the correction value.
Testing can be performed continuously.
For details, see
"5 Executing Continuous Tests" (page C-37) in "PART C Rockwell Hardness Test,
Plastic Hardness Test (ISO2039-2)".
Whenoperatingthespecimenstage,donotputyourhandsorngersundertheindenter.Theymaybecome
pinched.
1
Touch the home tab of the "condition setup" tab block.
» The home screen is displayed.
Tips
Tests cannot be executed unless the home
screen is displayed.
2
Slowly turn the up/down handle clockwise until the automatic brake engages and the up/down
handleisxed.
Tips
Whentheup/downhandleentersthexedposition,removeyourhandfromtheup/downhandle.
» The test is executed.
Tips
The background color of the hardness value is yellow when you are performing a test to obtain the correction
value.
Standard screen Simple screen