No. 99MBG518A
G-16
PART G : 2 Test Conditions
2.1.4 Plastic Hardness Test (ASTM D785-B)
No. Name Details
①
Scale button Thescaleisxedto[a–p].
②
Indenter selection Theballindenterisxedto1/2"(diameter:12.7mm).
③
Correction value If the correction value is set, the lamp at the top of the test force
lights up in green. If the correction value is not set, the lamp at the
top of the test force lights up in red.
A test can be correctly performed when the lamp at the top of the
test force lights up green.
The background color for the hardness value on the home screen is
as follows:
y White: A test can be run.
y Yellow: The state is either pre-correction or during correction.
④
[import] button Imports the results displayed on the home screen as the correction
value.
⑤
[Change] button Enters the correction value directly.
⑥
[Clear] button Clears the correction value.
⑦
Test force selection Thetestforceisxedat60(588.4N).
If the correction value is set, the lamp at the top of the test force
lights up in green. If the correction value is not set, the lamp at the
top of the test force lights up in red.
A test can be correctly performed when the lamp at the top of the
test force lights up green.