PART E : 3 Method to Set a Group and Test Multiple Points
3.8
Testing the Test Block
Performtestswithtestblocks(threeormoretypes)withdierenthardnessvalues.
3.8.1 Setting the Test Block
Set the test block on the specimen stage.
For details, see
"2.7.1 Setting the Test Block" (page E-10).
3.8.2 Executing the Test
Execute the test.
Whenoperatingthespecimenstage,donotputyourhandsorngersundertheindenter.Theymaybecome
pinched.
1
Touch the home tab of the "condition setup" tab block.
» The home screen is displayed.
Tips
Tests cannot be executed unless the home
screen is displayed.
2
Slowly turn the up/down handle clockwise until the automatic brake engages and the up/down
handleisxed.
Tips
Whentheup/downhandleentersthexedposition,removeyourhandfromtheup/downhandle.
» The test is executed.
Tips
The background color of the hardness value is yellow when you are performing a test to obtain the correction
value.