No. 99MBG518A
G-30
PART G : 3 Execution Conditions
3.3
External Output
Set the output method to external devices (USB output, serial output, SPC output, and USB memory).
The output settings to each external device can be set to either OFF, manual, or automatic.
If the output settings are set to OFF, data is not output even if an external device is connected.
If the output settings are set to automatic, the results are output each time a test is run.
If the output settings are set to manual or automatic, touch the hardness display after a test to output the
data.
External output from the hardness display of the home screen (manual settings)
Home screen External output method
Standard screen Touch the hardness display.
Simple screen Touch the hardness display.
List average screen Print button
List screen Touch the hardness display: Last test results
Print button: All test results
Multi-test screen Output cannot be performed from the home screen.
Even if output is enabled (manual or automatic), the results may not be output correctly if a USB device is not
connected.
For details about the DP-1VA LOGGER printer connection, see "1 Connecting with Printer DP-1VA
LOGGER" (page K-1) in "PART K Supplement".
For details about the external output, see
"2 External Output" (page K-3) in "PART K Supplement".
No. Name Details
①
USB output settings SpeciesthesettingsforoutputtingdatatoaUSBterminal.
With USB output, the average values and dispersion are output.
②
Serial output settings Speciesthedataoutputsettingstoaserialterminal.
With serial output, the average values and dispersion are output.
③
SPC output settings SpeciesthedataoutputsettingstoanSPCterminal.
Only the test data is output to the SPC.
④
USB memory output
settings
SpeciesthesettingsforoutputtingdatatoUSBmemory.
Only the test data is output to the USB memory.