No. 99MBG518A
G-3
PART G : 1 Home Screen
1.1
Standard Screen
This screen is the general display screen. This screen displays the hardness value, as well as informa-
tion such as the current number of samples, the conversion value, GO/NG judgment, the scale that is
set, and the time settings.
No. Name Details
Hardness value Displays the test results.
If external output is set to a value other than [OFF], press and hold to
perform external output.
For details about external output, see "3.3 External Output" (page
G-30).
Scale Displays the scale that is currently set (indenter/test force).
To change the test method, see "5.2 Test Method" (page G-50).
To change the scale, see "2.1 Scale" (page G-13).
[Continuous] button Touching the [Continuous] button activates the display and enables
testing.
For details, see "5 Executing Continuous Tests" (page C-37) in
"PART C Rockwell Hardness Test, Plastic Hardness Test (ISO2039-
2)".
[Run] button When continuous testing is enabled, this button can be used for sub-
sequent points. Touch the [Run] button to start a test.
Time settings Displays the preliminary test force retention duration F0, the total test
force retention duration F, and the duration until reading after unload-
ing the total force that are set.
For details, see "2.2 Test Force/Time" (page G-18).
Conversion Displays the conversion results and conversion scale for the test
results.
For details, see "3.1 Conversion Settings" (page G-27).
GO/NG judgment Displays GO/+NG/–NG for the test results.
For details, see "3.2.1 GO/NG Judgment" (page G-28).
Sample No./group Ifgroupmeasurementiso󰀨,displaysthecurrentnumberoftested
points and maximum number of points.
If group measurement is on, displays the current number of tested
points/set number of points and the number of groups.
For details, see "3.2.2 Group" (page G-29).
Status icon Displays the currently set functions as icons.
[Setting] button Displays the condition number that is currently set.
For details, see "5.6.1 System Conditions" (page G-56).

Table of Contents