No. 99MBG518A
C-9
PART C : 2 Method to Test One Point
2.4
Mounting the Indenter
Mount the indenter set in "2.3 Scale Settings" to the main unit of the testing machine.
For details about mounting the indenter, see
"3 Inspection and Replacement of Consumable Parts"
(page H-5) in "PART H Inspection and Maintenance".
2.5
Test Time Settings
Set the test time.
1
Touch the [Force/Time] button on the "basic operation" block.
» The Test Time settings screen is dis-
played.
2
Set the conditions.
For details about various conditions, see
"2.2 Test Force/Time" (page G-18) in "PART
G Settings".
2.6
Correction Value Settings
Set a correction value as necessary.
Set it in advance when, for example, a specimen with a curved surface (specimen such as a cylindrical
shape) or an error occurs in test results between testing machines or operators.
For details about correction values, see
"2.3 Correction" (page G-23) in "PART G Settings".