No. 99MBG518A
G-11
PART G : 1 Home Screen
1.5
Multi-Test Screen (When the Fine-Adjustment
Table for Jominy Testing is Connected)
No. Name Details
①
Scale Displays the scale that is currently set (test force).
To change the test method, see "5.2 Test Method" (page G-50).
②
List display Displays the test results in a list.
From left to right, the list displays the test number (1–10240), X (axis
[mm]), Y (axis), and the hardness judgment (GO/+NG/–NG).
X(axis)displaysthetestpositionsforwhenthene-adjustmenttable
forJominytestingisconnected.Thebackgroundofthedataeld
thatisselectedisdisplayedinblue.Touchadierenteldtoselect
thatdataeld.
③
Up/down buttons Touch either button to move the line selection up or down.
The list also scrolls if data is hidden.
④
Status icon Displays the currently set functions as icons.
⑤
Counter display Displays the position of the X-axis. If the axis is moved to the coordi-
nates that are selected in the list display, the value blinks and a test
can be run.
⑥
[Continuous] button Touching the [Continuous] button activates the display and enables
testing. For details, see "5 Executing Continuous Tests" (page
C-37) in "PART C Rockwell Hardness Test, Plastic Hardness Test
(ISO2039-2)".
⑦
[Run] button When continuous testing is enabled, this button can be used for
subsequent points. Touch the [Run] button to start a test.
⑧
Position input button Used to enter the test position for the X-axis.
⑨
Clear button Resets only the results displayed for all of the hardness values to
0.0.
Afterresetting,ifyouspecifyaspecicnumberandrunatest,only
the number that is selected will be retested.
⑩
[Setting] button Displays the condition number that is currently set.
For details, see "5.6.1 System Conditions" (page G-56).