No. 99MBG518A
B-19
PART B : 6 Installing or Removing the Bellows
6.2
Installing the Bellows
1
Lower the specimen stage.
1 Turn the up/down handle counterclock-
wise to lower the specimen stage until
the gap between the indenter and the
vertical shaft is at least 100 mm.
2 Remove the indenter.
For details about removal, see
"3.1 Replacing the Indenter" (page
H-5) in "PART H Inspection and Main-
tenance".
3 Remove the specimen.
4 Remove the anvil.
No. Name
Indenter
Vertical shaft
Specimen
Anvil
2
Turn the inner rings (x2) counterclockwise to remove them.
No. Name
Inner rings (x2)
Vertical shaft