No. 99MBG518A
F-25
PART F : 3 Method to Set a Group and Test Multiple Points
3.10.5 Executing the Test (Second Point)
Execute the second test.
Whenoperatingthespecimenstage,donotputyourhandsorngersundertheindenter.Theymaybecome
pinched.
Tips
Tests cannot be executed unless the home screen is displayed.
1
Set the specimen.
For details, see "3.9 Setting the Specimen" (page F-21).
2
Slowly turn the up/down handle clockwise until the automatic brake engages and the up/down
handleisxed.
Tips
Whentheup/downhandleentersthexedposition,removeyourhandfromtheup/downhandle.
» The test is executed.
3