No. 99MBG518A
D-26
PART D : 3 Method to Set a Group and Test Multiple Points
3.8
Test for the Specimen for Obtaining the Correction
Value
Execute the test with the specimen for obtaining the correction value.
3.8.1 Setting the Specimen for Obtaining the Correction Value
Set the specimen for obtaining the correction value on the specimen stage.
For details, see
"2.7.1 Setting the Specimen for Obtaining the Correction Value" (page D-10).
3.8.2 Executing the Test
Execute the test with the specimen for obtaining the correction value.
CAUTION
Whenoperatingthespecimenstage,donotputyourhandsorngersundertheindenter.Theymaybecome
pinched.
1
Touch the home tab of the "condition setup" tab block.
» The home screen is displayed.
Tips
Tests cannot be executed unless the home
screen is displayed.
2
Slowly turn the up/down handle clockwise until the automatic brake engages and the up/down
handleisxed.
Tips
Whentheup/downhandleentersthexedposition,removeyourhandfromtheup/downhandle.
» The test is executed.
Tips
The background color of the hardness value is yellow when you are performing a test to obtain the correction
value.