No. 99MBG518A
D-28
PART D : 3 Method to Set a Group and Test Multiple Points
3.9
Correction Value Setup
Set up the test result (correction value) of the specimen for obtaining the correction value.
1
Touch the [Test] tab of the "condition setup" tab block.
» The Test settings screen is displayed.
List average screen List screen
2
Touch the [import] button.
» The correction value will be set up. (The
lamp above the test force will switch
from red to green.)
Tips
You can also enter the correction value by
touching [Change] described in the main
text.