No. 99MBG518A
G-20
PART G : 2 Test Conditions
2.2.3 Plastic Hardness Test (ISO 2039-2)
No. Name Details
①
Force Displays the preliminary test force and total test force. (The setting cannot
be set directly in this screen.)
The setting can be set in the scale settings screen.
②
Time Therearevetypesoftimesettingsavailable.
For the range of settings, see "■Relationshipbetweenthetestcycle
and time".
■Relationship between the test cycle and time
1 2
3
4 5
t
Zero set
Read
Display name Description Setting range
1 Load F0 Touching specimen to preliminary test force loading time 1–2 sec.
2 Duration F0 Preliminary test force retention time 1–120 sec.
3 Load F After preliminary test force retention to total test force
loading time
2–10 sec.
4 LoadDurat. After preliminary test force retention to total test force
retention completion time
3–120 sec.
5 UnloadRead After total test force retention to data read 3–120 sec.