No. 99MBG518A
G-54
PART G : 5 System Conditions
5.5
System
No. Name Details
Version Displaysversioninformationwhenthe[Conrm]buttonistouched.
When contacting Mitutoyo, note the displayed version beforehand.
Loading When testing is performed, the number of times is counted and the
display is incremented by 1 every time testing is performed.
Touchingthe[Clear]buttondisplaysaconrmationmessage.To
clear the count, touch the [OK] button; otherwise touch the [Cancel]
button.
The count can be used as an approximate guide to indenter replace-
ment, maintenance period, etc.
Language Changes the displayed language.
Displays the currently set language.
Note that if the language is set to an unknown one, it may not be
possible to return to the previous language.
Time Sets the date and time.
For details, see "5.5.1 Time" (page G-55).
[MAINTENANCE]
button
Selectwhenperformingvericationoflength(depth)andtestforce.
Whenperformingverication,loadcell,indexmaster,orotherjigs
will be required.