EasyManua.ls Logo

Mitutoyo HR-530 Series - System Conditions Configuration; Guide Function

Mitutoyo HR-530 Series
356 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
No. 99MBG518A
G-49
PART G : 5 System Conditions
5
System Conditions
Set the system conditions.
The setup screen for system conditions is displayed by touching the [System] tab on "condi-
tion setup" tab block.
[Guide], [Test Method], [Display], [Equipment], [System], and [Settings] are shown at the bot-
tom of the screen.
The [Test Method] screen is displayed right after this screen is displayed.
5.1
Guide
This function derives the relationship between the minimum thickness and hardness of the specimen
duringRockwellhardnesstestandRockwellsupercialhardnesstest.
To derive the minimum thickness, touch the [Min. thick.] button in [Targ. Value] to activate it. When the
[Hardness] button in [Input Item] is touched and hardness is input, the minimum thickness is derived.
To derive the hardness, touch the [Hardness] button of [Targ. Value] to activate. When [Min. thick.] in
[Input Item] is touched and minimum thickness is input, the hardness is derived.
Tips
UsableonlyintheRockwellhardnesstestandRockwellsupercialhardnesstest.
The relationship of applicable ranges is calculated from the graphs of JIS Z 2245 Annex B or ISO 6508-1
Appendix B.