PART G : 2 Test Conditions
2.2
Test Force/Time
2.2.1 Rockwell/RockwellSupercialHardnessTest
No. Name Details
Force Displays the preliminary test force and total test force. (The setting cannot
be set directly in this screen.)
The setting can be set in the scale settings screen.
Time Speciesthevarioustimesettings.
There are six types of time settings available.
For the range of settings, see "■Relationshipbetweenthetestcycle
and time".
■Relationship between the test cycle and time
F
t
1 2 3 4 5 6
Read
Zero set
Display name Description Setting range
1 Load F0 Touching specimen to preliminary test force loading time 1–2 sec.
2 Duration F0 Preliminary test force retention time 1–120 sec.
3 Load F After preliminary test force retention to total test force
loading time
2–10 sec.
4 Duration F Total test force retention time 1–120 sec.
5 Unload F After total test force retention to test force unloading time 110 sec.
6 Read Preliminary test force retention to data read 1–120 sec.