No. 99MBG518A
G-10
PART G : 1 Home Screen
1.4
List Screen
This screen displays the test results for the current test and four previous tests. This is a simple multi-
test screen to display multiple test results, the average and dispersion of those test results on a single
screen.
No. Name Details
①
Hardness value Displays test results.
If external output is set to a value other than [OFF], press and hold to
perform external output.
For details about external output, see "3.3 External Output" (page
G-30).
②
Scale Displays the scale that is currently set (indenter/test force).
To change the test method, see "5.2 Test Method" (page G-50).
To change the scale, see "2.1 Scale" (page G-13).
③
Previous test results y Displays the test results from the four most recent tests on the home
screen.
y Oldertestresultscanbeconrmedontheresultsscreen.
y For details, see "4 Results" (page G-35).
④
Dispersion and
average
Displays the dispersion (R = Max. – Min.) and average value (Ave.) for
theresultsdisplayedonthescreen(uptovetests)orforonegroup.
⑤
[Continuous] button Touching the [Continuous] button activates the display and enables
testing. For details, see "5 Executing Continuous Tests" (page
C-37) in "PART C Rockwell Hardness Test, Plastic Hardness Test
(ISO2039-2)".
⑥
[Run] button When continuous testing is enabled, this button can be used for subse-
quent points. Touch the [Run] button to start a test.
⑦
Clear button Touch this button to delete the [Hardness] currently displayed.
⑧
Data delete button Deletes the hardness value displayed on the home screen.
: Deletes the results displayed on the screen (up to
vetests)ortheresultsforonegroup.
: Touch a previous test result to select it.
Touch this button to delete the test result that is selected.
⑨
Print button Touch this button to output the test results to an external device.
Tips
Test results cannot be output to the [SPC] external output.
⑩
GO/NG judgment Displays GO/+NG/–NG for the test results.
For details, see "3.2.1 GO/NG Judgment" (page G-28).
⑪
[Setting] button Displays the condition number that is currently set.
For details, see "5.6.1 System Conditions" (page G-56).