No. 99MBG518A
H-5
PART H : 3 Inspection and Replacement of Consumable Parts
3
Inspection and Replacement of
Consumable Parts
Parts such as the indenter are consumable parts and need to be replaced. Replace the con-
sumables by following the procedures below.
3.1
Replacing the Indenter
When replacing the indenter during a test, switch the screen of the display unit to the screen for test conditions,
and then replace the indenter. The testing machine may start a test and cause injury if the replacement is per-
formed while the display unit displays the home screen or the details screen for test results.
• Be careful that the indenter does not touch other parts such as the specimen or anvil during replacement.
• Always keep the tip of the indenter clean.
Tips
If an indent has been added to a dirty specimen or if an indent is added to a clean specimen after the indenter
hasbeentouchedbyanger,testingmaybedicultduetotheadhesionofdust,oil,orotherdirtaroundthe
indent.Inthiscase,lightlywipethetipoftheindenterwithacleancloth.Ifitisoily,wipeooilwithacleaner
such as alcohol.
1
Turn the up/down handle counterclockwise to lower the specimen stage.
Tips
Whenthereisaspecimenontheanvil,ensuresucientspacesothattheindenterdoesnottouchthespeci-
men.