No. 99MBG518A
B-1
PART B : 1 Installation Environment
1
Installation Environment
Use this product in the following installation environment to fully utilize its functions and
performance.
1.1
Temperature Range
y Vericationandcalibrationtemperature:23°C±5°C
y Usage temperature: 5 °C through 40 °C
y Storage temperature: -10 °C through 50 °C
1.2
Humidity Range
80 %RH or less and with no condensation
Tips
Thehumiditydoesnotdirectlyinuencetheaccuracyofthisproduct,butifthehumidityistoohigh,rustoccurs
moreeasilyonimportantpartsofthedevice,adverselyaectingelectricalcomponents.Therecommended
humidity is from 55 % through 65 %.
1.3
Vibration
Installthisproductinanareasubjecttominimalvibrationfromtheoorsuchasthosebelow.
If this product is installed in an area subject to vibrations, the testing machine may not obtain accurate test
results due to the transmission of the vibrations.
• Buildingsthathaveasolidfoundationwithooringwhichdoesnottransmitvibrationsfromothersources
• Areassubjecttominimalfoottrac
• Areas away from machinery that produce vibrations
• Therstoorofbuildings
• Buildingsawayfromroadswithalargeamountofvehicletrac
• Buildings without indoor cranes, vibration testing machines, or other similar devices
Vibration conditions Use anti-vibration measures when using this product in an area that exceeds
the conditions below.
ContactthenearestMitutoyosalesoceaboutanti-vibrationmeasures.
y Acceleration 0.5 Gal or less
y Frequency 5 Hz or less