No. 99MBG518A
G-17
PART G : 2 Test Conditions
2.1.5 Indent Depth Brinell Hardness Test (VDI/VDE2616)
No. Name Details
Scale button The scale cannot be set directly. The end of the scale varies accord-
ing to the material for the ball indenter of the "indenter selection".
Indenter selection Selects the material of the ball for the indenter.
W: Carbide ball S: Steel ball
Thediameteroftheballisxedto2.5mm.
Test force selection Selects the total test force. The test force switches to the test force
of the button that is touched.
If the correction value is set, the lamp at the top of the test force
lights up in green. If the correction value is not set, the lamp at the
top of the test force lights up in red.
A test can be correctly performed when the lamp at the top of the
test force lights up green.
[Setup] button Selects the correction value to use in a test.