PART E : 3 Method to Set a Group and Test Multiple Points
4
Set the reference value of each test result.
1 If necessary, set the indenter number.
2 Select the test result.
3 Touch the [Change Ref.] button.
» The [Ref.] dialog box is displayed.
4 Input the reference value (test block
value).
5 Touch the [Enter] button.
6 Repeat steps 2 to 4.
5
Set up the correction value.
1 Touch the [Setup] button.
» Theconrmationdialogboxisdisplayed.
2 Touch the [OK] button.
» The lamp above the [Setup] button
switches from red to green.
Tips
The correction value can also be saved to
USB memory.