No. 99MBG518A
G-25
PART G : 2 Test Conditions
●
Multi-point correction setup
No. Name Details
[Indenter] Speciestheinputfortheindenternumber.
Scale display Displays the scale that is currently set.
List display Displays [No], [Ref.], and [Hardness] in a list.
The background of the line that is selected is displayed in blue.
Touch a line to select the line.
Up/down buttons Touch either button to move the line selection up or down.
In addition, the list also scrolls if data is hidden in the list display.
[import] button Imports the test results that are saved and displays that data in the
[Hardness]eld.
[Add.] button Adds correction points.
[Change Ref.] button Changes the reference values that are selected.
[Change HD] button Changes the hardness values that are selected.
[Setup] button Creates a quadratic equation based on the data for correction.
The lamp at the top of the button lights up green to indicate success.
The lamp lights up red to indicate failure of the calculation. If the
lamp is red, correction cannot be performed for the test results.
[Clear] button Clears correction.
Correction value read
button
Loads the correction values from the USB memory.
Correction value save
button
Saves the correction values that are set to the USB memory.
[OK] button Closes the multi-point correction setup screen and returns to the
correction screen.
■Shift correction
If shift correction is not selected, touch the [Shift] button.
Touch the [Shift] button to display the numerical keys. Enter the correction value.