EasyManua.ls Logo

Planmeca PlanMill 40 - Page 56

Planmeca PlanMill 40
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Tool changer calibration Chapter C: Calibration
54 Planmeca PlanMill 40 Technical manual
Figure 46: Insert calibration pin in left collet
13. Insert the calibration pin in the right collet, hold in place and touch the Auto Both
button (collet closes).
NOTE
When holding the calibration pin, the system automatically inserts it to the collet. Check that
the calibration pin is correctly inserted to the collet.
Figure 47: Insert calibration pin in right collet
14. Enter the Tool Holder Calibration Fixture number in the Fixture # field the touch the
Auto Both button.
Enter all digits of the Tool Holder Calibration Fixture number to the Fixture # field.

Table of Contents

Related product manuals