EasyManua.ls Logo

Planmeca PlanMill 40 - Page 67

Planmeca PlanMill 40
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter C: Calibration 4 Origin calibration
Technical manual Planmeca PlanMill 40 65
Figure 61: Disconnect Calibrator cables and remove Origin Calibration Fixture
17. The mill will start profiling the calibration pins. The lid opens and the message will
appear: "Remove Calibration Pins. Finally, close lid, and touch Origin button. Wait for
re-initialization to complete (as indicated by green LED’s) after touching the button."
Follow these instructions.
Figure 62: Remove Calibration Pins
18. Message will appear: "Calibration is complete. Wait for re-initialization to complete (as
indicated by green LED’s) then mill the Refine Calibration job, and perform refine_2
calibration."

Table of Contents

Related product manuals