ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-11
No. Message Interpretation Operator action
529 Inlet LS failed to
calibrate
Inlet liquid sensor failed to
calibrate.
- Repeat the liquid
sensor calibration.
- Contact Radiometer
service representative.
531 Sensors LS failed
to calibrate
Liquid sensor near the sensor
cassette failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace if
necessary.
- Contact Radiometer
service representative.
537 OXI LS failed to
calibrate
OXI module liquid sensor
failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
581 OXI spectrum
mismatch
Spectrum deviates from the
expected blood or QC
spectrum. Measurement may
be unreliable.
- Check the patient
record for medication
containing possible
interfering substances.
- Start a calibration.
- Contact Radiometer
service representative.
582 tHb calibration
cuvette factor
outside limits
tHb calibration failed. - Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- Successful tHb
calibration.
584 tHb calibration
wavelength
outside limits
tHb calibration failed. - Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- A successful tHb
calibration
588 Measured QC
value lower than
statistical range
he parameter value is below
the lower limit of the user-
defined statistical range.
Measurement is not included
in statistics.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference manual
for details on the
evaluation of the
results.