ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-11
No. Message Interpretation Operator action
529 Inlet LS failed to
calibrate
Inlet liquid sensor failed to
calibrate.
- Repeat the liquid
sensor calibration.
- Contact Radiometer
service representative.
531 Sensors LS failed
to calibrate
Liquid sensor near the sensor
cassette failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace if
necessary.
- Contact Radiometer
service representative.
537 OXI LS failed to
calibrate
OXI module liquid sensor
failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
581 OXI spectrum
mismatch
Spectrum deviates from the
expected blood or QC
spectrum. Measurement may
be unreliable.
- Check the patient
record for medication
containing possible
interfering substances.
- Start a calibration.
- Contact Radiometer
service representative.
582 tHb calibration
cuvette factor
outside limits
tHb calibration failed. - Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- Successful tHb
calibration.
584 tHb calibration
wavelength
outside limits
tHb calibration failed. - Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- A successful tHb
calibration
588 Measured QC
value lower than
statistical range
T
he parameter value is below
the lower limit of the user-
defined statistical range.
Measurement is not included
in statistics.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference manual
for details on the
evaluation of the
results.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
TypeBlood Gas Analyzer
Analysis Time35 seconds
Parameters MeasuredpH, pCO2, pO2
Sample Volume65 μL
ConnectivityEthernet, USB
CalibrationAutomatic
Data ManagementIntegrated data management system
Sample TypeWhole blood, plasma, serum
Power Requirements100-240 V AC, 50-60 Hz, 150VA
Operating Temperature15-32 °C (59-90 °F)