10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-12
No. Message Interpretation Operator action
589 Measured QC
value higher than
statistical range
The parameter value is
above the upper limit of the
user-defined statistical
range. Measurement not
included into statistics.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference manual
for details on the
evaluation of the
results.
593 Insufficient
sample
Sample volume is too small
for the selected measuring
mode. Affected parameters
will be marked with "?".
- Repeat the
measurement, ensuring
sufficient sample
volume.
- Contact Radiometer
service representative.
595 Liquid sensor
calibration error
One or more of the liquid
sensors failed calibration.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
606 Cal expired (pH) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful calibration.
608 Cal expired
(pCO
2
)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
609 Cal expired (pO
2
)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
610 Cal expired (K) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
611 Cal expired (Na) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
612 Cal expired (Ca) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.