ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-13
No. Message Interpretation Operator action
613 Cal expired (Cl) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
614 Cal expired (Glu)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
615 Cal expired (Lac)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as “…..”.
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
616 Cal expired (OXI)
Too long time elapsed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 1- or 2-
point calibration.
641 ABL/DMS PC
restarted
The analyzer was restarted
from power off.
- No action required.
For information only.
642 ABL/DMS PC
connected to wet
section
Added by DMS PC when
connection to the wet section
is obtained.
- No action required.
643 ABL/DMS PC
disconnected
from wet section
he connection from the DMS
PC to the wet section is lost.
- Shut down and restart
the analyzer.
- Contact Radiometer
service representative.
648 Calibration failed
or not accepted
The last calibration was
aborted or not accepted.
- Check solution pack
status and replace, if
necessary.
- Check sensor cassette
status and replace, if
necessary.
- Check for and remedy
system messages.
- Repeat the calibration.
Removal condition:
- Successful calibration.
662 Barometer out of
range
Measured barometer value is
outside the measuring range:
60-106.7 kPa.
- Contact Radiometer
service representative.
669 QC value outside
control range
Measured parameter value is
outside control range.
- Verify the procedure
and repeat
measurement.
- Refer to Quality
Control Systems
Reference Manual.