10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-14
No. Message Interpretation Operator action
679 Barometer error The measured parameter
may be unreliable due to
barometer error.
- Contact Radiometer
service representative.
682 OXI module not
active
The OXI module is not
responding due to an internal
communication problem, or
the software configuration
does not match the analyzer
type.
- Shut down the
analyzer, using the
Temporary Shutdown
function; then restart it.
- Contact Radiometer
service representative.
Removal condition:
- OXI module ready, or
software configured
without OXI module
support.
688 ctHb/ceHb too
low for OXI
calculation
ctHb < 1 mmol/L, or ceHb <
0.75 mmol/L. If ctHb is too
low, FHHb, FO
2
Hb, FCOHb
and FMetHb are not
calculated. If ceHb = cHHb +
cO
2
Hb is too low, sO
2
is not
calculated.
- If Oxi derivates are
wanted, elevate tHb
and/or sO
2
.
692 ABL not
connected to
RADIANCE
The analyzer is not
connected to RADIANCE.
- Contact your
RADIANCE/IT engineer.
- Check RADIANCE
Communication Setup
including TCP/IP
address, port no. and
password.
- Check that RADIANCE
is responding.
- Check network
connections.
Removal condition:
- RADIANCE connection
established or disabled.
693 ABL not
connected to
RADIANCE -
incorrect
password
The analyzer was refused
connection to RADIANCE due
to incorrect password.
- Enter the correct
password in the
analyzer's RADIANCE
Communication Setup.
Removal condition:
- RADIANCE connection
established or disabled.
694 ABL connected to
RADIANCE
The analyzer is connected to
RADIANCE.
- No action required.
695 ABL disconnected
from RADIANCE
The analyzer was
disconnected from
RADIANCE.
- No action required.
696 ABL<>RADIANCE
communication
error
Communication error
between the analyzer and
RADIANCE.
- Contact Radiometer
service representative.