10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-16
No. Message Interpretation Operator action
734 General WSM
exception
The data management
system establishes
connection to the analyzing
unit, or the connection is
lost.
- Wait a few minutes for
the connection to
establish.
- Restart the analyzer.
- If the error persists,
contact Radiometer
service representative.
745 Low disk space Free disk space is low. - Move archive files to
another storage device.
Removal condition:
- Sufficient free hard
disk space.
766 ABL not
connected to
RADIANCE - no
RADIANCE
connection
license
The analyzer has been
refused connection to
RADIANCE because there is
no connection license
available on RADIANCE.
- Contact RADIANCE/IT
engineer or Radiometer
service representative.
Removal condition:
- Connection to
RADIANCE established.
767 ABL not
connected to
RADIANCE - ABL
StatLink version
too high
The analyzer has been
refused connection to
RADIANCE because the ABL
StatLink version is higher
than the RADIANCE StatLink
version.
- Contact RADIANCE/IT
engineer or Radiometer
service representative.
Removal condition:
- RADIANCE connection
established.
768 ABL not
connected to
RADIANCE - ABL
StatLink version
too low
The analyzer has been
refused connection to
RADIANCE because the ABL
StatLink version is lower
than the RADIANCE StatLink
version.
- Contact RADIANCE/IT
engineer or Radiometer
service representative.
Removal condition:
- RADIANCE connection
established.
769 ABL<>RADIANCE
communication
error - XML
packet could not
be parsed
Communication error
between the analyzer and
RADIANCE.
- Contact RADIANCE/IT
engineer or Radiometer
service representative.
770 Failed to restore
Custom Setup
The setup could not be
restored.
- Download the setup
data from another
floppy disk, hard disk or
network drive.
- Contact Radiometer
service representative if
the error persists.
771 Succeeded to
restore Custom
Setup
Restoring of setup is
completed.
- No action required.
772 User Activity: User activity logged by
operator.
- No action required.