ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-19
No. Message Interpretation Operator action
819 Lac locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
820 tHb locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
821 MetHb locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
822 COHb locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
823 HHb locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
824 O
2
Hb locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
825 sO
2
locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.