10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-18
No. Message Interpretation Operator action
811 pCO
2
locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
812 pO
2
locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
813 K locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
814 Na locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
Await corrective actions
initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
815 Cl locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
816 Ca locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
818 Glu locked The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.