ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-21
No. Message Interpretation Operator action
841 tHb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
842 MetHb unlocked The message is used in the
Activity Log to
indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
843 COHb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
844 HHb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
845 O
2
Hb unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
846 sO
2
unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
847 HbF unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
848 tBil unlocked The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
852 RADIANCE: Message from RADIANCE. - No action required. For
information only.
855 Base Excess out
of range
Base Excess exceeds the +/-
30 mmol/L range.
- For information only.
No analyzer error was
detected.
875 Sample aged
he specified limit for sample
age has been exceeded.
- Draw and analyze new
sample.
885 Cyclic QC
schedule reset
from RADIANCE
The cyclic QC schedule has
been reset and all related
reminders have been
removed as a result of a
RADIANCE command.
- No action required. For
information only.
886 LIS/HIS: No valid
POCT1A DML
Device ID file
A file with a valid Device ID
does not exist. A valid Device
ID is needed in order to use
the POCT1A DML protocol.
- Contact Radiometer
service representative
to obtain a Device ID
file.
Removal condition:
- Valid Device ID found.