10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-22
No. Message Interpretation Operator action
963 Leak currents
detected
Leak currents were detected
during system calibration and
may distort measuring
results.
- Replace liquid- or
sensor cassette.
- Contact Radiometer
service representative.
964 Leak currents
detected
Leak currents were detected
during system calibration and
may distort measuring
results.
- Check sensor cassette
status and replace, if
necessary.
- Contact Radiometer
service representative.
970 Replace solution
pack
This message is shown when
the solution pack needs to be
replaced. The analyzer will
enter "User-intervention
required".
- Replace solution pack.
971 Replace sensor
cassette
This message is shown when
the sensor cassette needs to
be replaced. The analyzer
will enter "User-intervention
required".
- Replace sensor
cassette.
972 Sensor
conditioning due
The sensor cassette must
soon be replaced. A new
sensor cassette should
preferably be conditioned
now.
- Start conditioning a
new sensor cassette.
973 Printer paper
must be replaced
No more paper in printer. - Insert new printer
paper.
978 Flow selector
calibration error
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
979 Inhomogeneous
rinse solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
980 Inhomogeneous
QC1 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
981 Inhomogeneous
QC2 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.