ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-23
No. Message Interpretation Operator action
982 Inhomogeneous
QC3 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
983 Inhomogeneous
cal 3 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
984 The analyzer
could not
aspirate
homogeneous
calibration
solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1000 Number of pO
2
hardware data
fail
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1001 Timeout while
waiting for pO
2
hardware data
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1002 pO
2
dark data is
out of range
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1004 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1005 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1006 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1007 Missing oxygen
calibration
No calibration data exists for
oxygen.
- Perform a calibration.
1008 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.