10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-24
No. Message Interpretation Operator action
1009 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1010 Oxi data
collection error
Oxi hardware problem - Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1011 Oxi has no Blank
Cal
Missing Blank Cal. Not
necessarily a hardware error.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
Removal conditioning:
- Successful Blank
calibration
1012 Oxi has no
sample spectrum
The system has not made a
sample measurement yet, or
there is a hardware problem.
- Repeat the
measurement.
- Restart the analyzer.
- Contact Radiometer
service representative.
1013 Oxi data
collection error
Oxi hardware error - Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1014 Oxi Blank Cal.
intensity too high
The spectrometer received
too high
light intensity during
Blank Cal.
- Check solution pack.
During Oxi Blank
calibration, the cuvette
must be filled with
liquid.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1015 Oxi sample
intensity too high
The spectrometer received
too high light intensity during
sample measurement.
- Check solution pack.
During Oxi Blank
calibration, the cuvette
must be filled with
liquid.
- Perform a calibration.
- Repeat the sample
measurement.
1016 Oxi Blank Cal.
intensity too low
The spectrometer received
too low light intensity during
Blank Cal.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1017 Oxi sample
intensity too low
The spectrometer received
too low light intensity during
sample measurement.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.