ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-25
No. Message Interpretation Operator action
1018 Oxi electronic
adjustment error
Oxi hardware problem. - Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1019 Oxi Blank Cal.
outside limits
Peak value of Blank Cal.
spectrum intensity is outside
acceptance limits.
- Check solution pack.
The cuvette must be
filled with liquid during
Blank calibration.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1020 Oxi neon
intensity outside
limits
Oxi hardware problem. - Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1021 Oxi neon
correction
outside limits
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1022 Oxi background
correction
outside limits
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1023 Oxi spectrometer
memory read
problem
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1024 Oxi spectrometer
memory write
problem
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1025 Oxi hemolyzer
tuning problem
Oxi hardware problem. - Restart the analyzer
- Contact Radiometer
service representative.
1026 Oxi hemolyzer
frequency
problem
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1027 Oxi hemolyzer
temperature
deviation too
high
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1028 Oxi neon voltage
outside limits
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1029 Oxi light source
voltage outside
limits
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.
1030 Oxi hemolyzer
voltage outside
limits
Oxi hardware problem. - Restart the analyzer.
- Contact Radiometer
service representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals