10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-32
No. Message Interpretation Operator action
1147 Inlet opened
during rinse
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1148 Inlet open during
calibration
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1149 Inlet open during
wet section
activity
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1150 Inlet closed
without
aspirating
sample
This message is shown in the
Activity Log when a
measurement has been
cancelled due to inlet being
closed before aspiration
could be completed.
- No action required. For
information only.
1151 Inlet not closed:
no sample
aspirated
T
his message is shown in the
Activity Log when a
measurement has been
cancelled due to inlet being
closed too late.
- No action required. For
information only.
1152 The solution pack
chip data could
not be read or
written during
replacement
T
his message is shown in the
Activity Log when a
replacement of the sensor
cassette or solution pack has
failed. The reason was that it
was impossible to
communicate with the chip
on the consumable.
- Repeat replacement
operation.
1157 No valid FTC
programs
detected
System error. - Contact Radiometer
service representative.
1165 Solution pack not
properly installed
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1166 Solution pack
expired
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals