ABL90 FLEX reference manual 10. Troubleshooting, updated for software version 2.3
10-31
No. Message Interpretation Operator action
1125 An unregistered
and used sensor
was installed
during
replacement
T
his message is shown in the
Activity Log after a sensor
cassette replacement. It
informs that the sensor
cassette installed is already
used and no information
exists about the conditioning
hereof.
- No action required. For
information only.
1126 A registered
sensor had been
used before
installation
T
his message is shown in the
Activity Log after a sensor
cassette replacement. It
informs that the sensor
cassette installed has been
used before.
- No action required. For
information only.
1134 The chip
information for
the solution pack
cannot be read
or written
T
his message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
- Restart the analyzer.
- Contact Radiometer
service representative.
1135 The solution pack
has been used
before
T
his message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
1140 The solution pack
has used the
maximum
number of
measurements at
installation
T
his message is shown in the
Activity Log after a failed
solution pack replacement.
- Reinstall the solution
pack.
1142 The printer door
is open. Printing
not possible
Printer door open. -
Ensure that the printer
paper is properly
installed.
- Close the printer door.
1143 Internal printer is
offline. Printing
not possible
Printer hardware error. -
Ensure that the printer
paper is properly
installed.
- Close the printer door.
1144 Check that
printer door is
closed and that
paper is present
Printer hardware error. -
Ensure that the printer
paper is properly
installed.
- Close the printer door.
1145 A printer error
has occurred.
Call service
technician
Printer hardware error. - Ensure that the printer
paper is properly
installed.
- Close the printer door.
1146 Printer paper
replaced
T
his message is shown in the
Activity Log after
replacement of printer paper.
- No action required. For
information only.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals