10. Troubleshooting, updated for software version 2.3 ABL90 FLEX reference manual
10-34
No. Message Interpretation Operator action
1175 Sensor
temperature
error
Hardware temperature error
(Thermistor).
- Ensure that the
ambient temperature is
between 15 and 32 °C.
- If the system has just
performed a cold start,
wait for the error to
disappear.
- Replace the fan filter,
if dirty.
- Shield the analyzer
from direct sunlight and
other heat sources.
- Contact Radiometer
service representative.
1176 A liquid sensor
error was
detected
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1177 A flow selector
error was
detected
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1178 A pump
calibration error
was detected
Shown on screen when
"User-intervention required"
has been entered due to this
reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1180 An error occurred
when trying to
communicate
with wet section
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1181 A software or
hardware error
exists in wet
section
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1183 Valve
malfunctioning
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL90 FLEX and is the answer not in the manual?

RADIOMETER ABL90 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL90 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals