Instrument Functions
R&S
®
FSVA/FSV
356Operating Manual 1176.7510.02 ─ 09
"Start"
The measurement is currently running.
"Stop"
The measurement has been stopped, or, in single sweep mode, the
end of the sweep has been reached.
Remote command:
ABORt on page 613
INITiate<n>:ESPectrum on page 758
Restore Standard Files ← Spectrum Emission Mask
Copies the XML files from the C:\R_S\instr\sem_backup folder to the C:
\R_S\instr\sem_std folder. Files of the same name are overwritten.
Remote command:
[SENSe:]ESPectrum:PRESet:RESTore on page 796
Result Evaluation
As a result of the Spectrum Emission Mask measurement, the measured signal levels,
the result of the limit check (mask monitoring) and the defined limit line are displayed in
a diagram (see also "Working with Limit Lines in SEM Measurements" on page 359).
Furthermore, the TX channel power "P" is indicated in relation to the defined power
class ranges.
Example:
For example, "P<31" is indicated if the lowest power class is defined from infinity to 31
and the power is currently 17 dBm.
In addition to the graphical results of the SEM measurement displayed in the diagram,
a result table is displayed to evaluate the limit check results (see also "Working with
Limit Lines in SEM Measurements" on page 359).
Measurement Functions