Remote Control
R&S
®
FSVA/FSV
834Operating Manual 1176.7510.02 ─ 09
Example: Sample setting:
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,
5us,MEAN,20
Performs a measurement sequence with the following settings:
Frequency = 935.2 MHz,
Resolution bandwidth = 1 MHz
Measurement time = 434 µs
Trigger source = VIDEO
Trigger threshold = 50 %
Trigger offset = 5 µs
Type of measurement = MEAN power
No. of measurements = 20
Sample query:
SENSe:MPOWer? 935.2MHz, 1MHz, 434us, VIDEO,
50PCT, 5us, MEAN, 20
Performs the same measurement and in addition returns the
results list immediately after completion of the last measure-
ment.
Sample result:
18.3,18.6,18.1,18.0,17.9,18.3,18.6,18.1,18.0,17.9,18.3,18.6,18.
1,18.0,17.9,18.3,18.6,18.1,18.0,17.9
SENSe:PMETer<p> Subsystem (Power Sensor, R&S FSV-K9)
This subsystem controls the device settings for power sensor measurements. It is only
available if the R&S FSV-K9 Power Sensor option is installed.
For details see Chapter 3.9, "Instrument Functions - Power Sensor (R&S FSV-K9)",
on page 538.
[SENSe:]PMETer<p>[:STATe].........................................................................................834
[SENSe:]PMETer<p>:DCYCle:STATe..............................................................................835
[SENSe:]PMETer<p>:DCYCle:VALue.............................................................................. 835
[SENSe:]PMETer<p>:FREQuency...................................................................................836
[SENSe:]PMETer<p>:FREQuency:LINK...........................................................................836
[SENSe:]PMETer<p>:MTIMe.......................................................................................... 836
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]...............................................................837
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt................................................................837
[SENSe:]PMETer<p>:ROFFset[:STATe]...........................................................................838
[SENSe:]PMETer<p>:TRIGger[:STATe]........................................................................... 838
[SENSe:]PMETer<p>:TRIGger:DTIMe..............................................................................839
[SENSe:]PMETer<p>:TRIGger:HOLDoff...........................................................................839
[SENSe:]PMETer<p>:TRIGger:HYSTeresis......................................................................839
[SENSe:]PMETer<p>:TRIGger:LEVel...............................................................................840
[SENSe:]PMETer<p>:TRIGger:SLOPe.............................................................................840
[SENSe:]PMETer<p>:UPDate[:STATe]............................................................................ 841
[SENSe:]PMETer<p>[:STATe] <State>
This command switches the power sensor measurement for the selected power sensor
on or off.
Remote Control – Commands