Remote Control
R&S
®
FSVA/FSV
936Operating Manual 1176.7510.02 ─ 09
Example:
HCOP:ITEM:WIND:TEXT 'comment'
Manual operation: See "Comment" on page 213
HCOPy:PAGE:ORIentation<1|2> <Orientation>
The command selects the format of the printout.
The command is only available if the output device is a printer (see HCOPy:
DESTination<1|2> on page 932).
Parameters:
<Orientation> LANDscape | PORTrait
*RST: PORT
Example:
HCOP:PAGE:ORI LAND
Usage: SCPI confirmed
Manual operation: See "Device Setup" on page 212
HCOPy:TDSTamp:STATe<1|2> <State>
This command includes or excludes the time and date in the printout.
The suffix <1|2> selects the printing device.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
HCOP:TDST:STAT OFF
Manual operation: See "Device Setup" on page 212
4.2.4.7 MMEMory Subsystem
The MMEMory (mass memory) subsystem provides commands which allow for access
to the storage media of the instrument and for storing and loading various instrument
settings.
The various drives can be addressed via the "mass storage unit specifier" <msus>
using the conventional DOS syntax. The internal hard disk is addressed by "C:". For
details on drives refer to Chapter 3.1.2, "Saving and Recalling Settings Files –
SAVE/RCL Key", on page 196.
The file names <file_name> are indicated as string parameters with the commands
being enclosed in quotation marks. They also comply with DOS conventions.
DOS file names consist of max. 8 ASCII characters and an extension of up to three
characters separated from the file name by a dot "." Both, the dot and the extension
are optional. The dot is not part of the file name. DOS file names do not distinguish
between uppercase and lowercase notation. All letters and digits are permitted as well
as the special characters "_", "^", "$", "~", "!", "#", "%", "&", "-", "{", "}", "(", ")", "@" and
Remote Control – Commands