Remote Control
R&S
®
FSVA/FSV
861Operating Manual 1176.7510.02 ─ 09
Manual operation: See "Auto Select" on page 255
4.2.3.16 SOURce Subsystem
The SOURce subsystem controls the output signals of the instrument.
● SOURce Subsystem commands...........................................................................861
● SOURce:POWer Subsystem................................................................................ 868
SOURce Subsystem commands
SOURce<n>:AM:STATe................................................................................................. 861
SOURce<n>:DM:STATe.................................................................................................862
SOURce<n>:EXTernal<generator>[:STATe]..................................................................... 862
SOURce<n>:EXTernal<generator>:FREQuency............................................................... 863
SOURce<n>:EXTernal<generator>:FREQuency:COUPling[:STATe]................................... 863
SOURce<n>:EXTernal<generator>:FREQuency[:FACTor]:DENominator.............................863
SOURce<n>:EXTernal<generator>:FREQuency[:FACTor]:NUMerator.................................864
SOURce<n>:EXTernal<generator>:FREQuency:OFFSet<m>.............................................865
SOURce<n>:EXTernal<generator>:FREQuency:SWEep[:STATe].......................................865
SOURce<n>:EXTernal<generator>:POWer[:LEVel]...........................................................866
SOURce<n>:EXTernal<generator>:ROSCillator[:SOURce]................................................ 866
SOURce<n>:FM:DEViation.............................................................................................867
SOURce<n>:FM:STATe................................................................................................. 867
SOURce<n>:FREQuency:OFFSet................................................................................... 868
SOURce<n>:AM:STATe <State>
This command turns external amplitude modulation of the tracking generator on and
off.
External I/Q modulation is switched off, if active. This command is available with a
tracking generator (option R&S FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SOUR2:AM:STAT ON
Switches on the external amplitude modulation of the tracking
generator for screen B.
Usage: SCPI confirmed
Manual operation: See "External AM" on page 478
See "Modulation OFF" on page 478