Instrument Functions
R&S
®
FSVA/FSV
390Operating Manual 1176.7510.02 ─ 09
Crest Crest factor (peak power – mean power)
0,01 % Level values over 0,01 % above mean power
0,1 % Level values over 0,1 % above mean power
1 % Level values over 1 % above mean power
10 % Level values over 10 % above mean power
Softkeys for APD Measurements
APD.............................................................................................................................390
Res BW.........................................................................................................390
# of Samples.................................................................................................391
Scaling.......................................................................................................... 391
x-Axis Ref Level..................................................................................391
x-Axis Range...................................................................................... 391
Range Log 100 dB....................................................................391
Range Log 50 dB......................................................................391
Range Log 10 dB......................................................................392
Range Log 5 dB........................................................................392
Range Log 1 dB........................................................................392
Range Log Manual................................................................... 392
Range Linear %........................................................................393
Range Lin. Unit.........................................................................393
y-Axis Max Value................................................................................393
y-Axis Min Value.................................................................................393
y-Unit % / Abs.....................................................................................393
Default Settings.................................................................................. 393
Adjust Settings....................................................................................394
Gated Trigger (On/Off)..................................................................................394
Gate Ranges.................................................................................................394
Adjust Settings..............................................................................................395
APD
Activates the function to measure the amplitude probability density (APD) and opens a
submenu.
For general information on calculating signal statistics see Chapter 3.3.5.8, "Calculat-
ing Signal Amplitude Statistics", on page 386.
Remote command:
CALCulate<n>:STATistics:APD[:STATe] on page 737
Res BW ← APD
Opens an edit dialog box to set the resolution bandwidth directly.
For correct measurement of the signal statistics the resolution bandwidth has to be
wider than the signal bandwidth in order to measure the actual peaks of the signal
amplitude correctly. In order not to influence the peak amplitudes the video bandwidth
is automatically set to 10 MHz. The sample detector is used for detecting the video
voltage.
Measurement Functions