Instrument Functions
R&S
®
FSVA/FSV
498Operating Manual 1176.7510.02 ─ 09
Figure 3-35: I/Q modulation
I/Q modulation is performed by means of the built-in quadrature modulator. The RF
signal is divided into two orthogonal I and Q components (in phase and quadrature
phase). Amplitude and phase are controlled in each path by the I and Q modulation
signal. By adding the two components an RF output signal is generated that can be
controlled in amplitude and phase.
Remote command:
SOURce<n>:DM:STATe on page 862
3.6.5 Displayed Information and Errors
Diagram header
In Tracking Generator measurement mode, some additional information is displayed in
the diagram header.
Label Description
INT TG: <source power> Internal tracking generator active
INT TG: <start power>… <stop power> Internal tracking generator with power sweep active
EXT TG <1|2>: <source power> External tracking generator (1 or 2) active
LVL Power Offset (see Chapter 3.6.3, "Configuring Tracking Gener-
ators", on page 483
FRQ Frequency Offset (see Chapter 3.6.3, "Configuring Tracking
Generators", on page 483
Measurement accuracy levels
NOR Normalization on;
No difference between reference setting and measurement
Instrument Functions – Tracking Generator (Options R&S FSV-B9/ R&S FSV-B10)