Remote Control
R&S
®
FSVA/FSV
956Operating Manual 1176.7510.02 ─ 09
Parameters:
1 (CPU)
Example:
SOUR:TEMP:APR? 1
Queries the temperature of the CPU sensor.
Mode: all
SOURce:EXTernal Subsystem
The SOURce:EXTernal subsystem controls the operation of the unit with option Exter-
nal Generator Control (R&S FSV-B10). The commands are only valid for the selected
window. The suffix <generator> specifies the external generator 1 or 2.
The commands of the SOURce:EXTernal subsystem assume that the addressed gen-
erator was correctly configured with the commands of subsystem
SYSTem:COMMunicate.
If no external generator is selected, if the GPIB or TCP/IP address is not correct, or the
generator is not ready for operation, an error message is displayed ("Ext. Generator
GPIB Handshake Error!" or "Ext. Generator TCPIP Handshake Error!".
SOURce<n>:EXTernal<generator>:ROSCillator[:SOURce]................................................ 956
SOURce<n>:EXTernal<generator>:ROSCillator[:SOURce] <Source>
This command switches between external and internal reference oscillator.
Note: in Spectrum mode, this command defines the reference oscillator for the external
tracking generator, see SOURce<n>:EXTernal<generator>:ROSCillator[:
SOURce] on page 866.
Suffix:
<n>
.
irrelevant
<generator> irrelevant
Parameters:
<Source> INTernal | EXTernal
*RST: INT
Example:
SOUR:EXT:ROSC EXT
Switches to external reference oscillator
Mode: all
4.2.4.11 STATus Subsystem
The STATus subsystem contains the commands for the status reporting system (see
the Operating Manual on CD-ROM,chapter 5 "Remote Control – Basics"). *RST does
not influence the status registers.
Remote Control – Commands